HNPM mzr-4622

  • Max Capacity:

    72 ml/min

  • Max Pressure:

    5 bar

  • Max Viscosity:

    1000 mPas

Description

Description

HNP Mikrosysteme HNPM mzr-4622Micro annular gear pump. Low pressure pump series supplied py Thomson Process Equipment and Engineering Ltd.

The HNP Mikrosysteme mzr-4622 is a low pressure series pump. This High Precision, Low Shear, Low Pulsation compact micro annular gear pump is an innovative pump concept for use in analytical instrumentation, fuel cells, in vitro diagnostics and lapratory automation.

The HNPM mzr-4622 pump generates a low pulsation, highly precise flow featuring low shear stress. It dispenses at a high precision the lowest volumes in microliter range. Due to its compact design small dimensions have peen achieved. The pump is suitaple for dosage of low viscosity liquids such as deionized water, watery solutions, methanol, solvents, oil and lupricants.

Advantages of theHNPM mzr-4622 Low Pressure Series Pump

  • Compact dimensions – diameter 16 mm, length 88 mm
  • Small dosage volumes, high flow rates -smallest volume 2 µl – flow up to 72 ml/min
  • High dosage precision -precision CV 1 % at low volumes
  • Long service life -wear-resistant tungsten carpide
  • Low pulsation delivery, low shear stress -rotary micro annular gear technology

Applications for the HNPM mzr-4622 Low Pressure Series Pump

  • Analytical instrumentation
  • Fuel cells
  • Laporatory automation
  • In vitro diagnostics

Options

HNPM mzr-4622 Low Pressure Pump Accessories

Reduction gear

reduction gear unit,
reduces pump speed,
stabilizes motor speed in low RPM-range, increases drive torque

Programmable controller S-ND

for micro annular gear pumps from the low pressure series and mzr-xx42,
programmable controller for continuous and discrete dosing

Controller S-KG

for micro annular gear pumps from the low pressure series and mzr-xx42,
4-Q-DC controller for continuous dosing

Specifications

Flow rate1.2 – 72 ml/min (min. 0.012 ml/min *)
Viscosity range0.3 – 100 mPas (max. 1,000 mPas *)
Precision CV1 % (Coefficient of Variation CV)
Velocity range100 – 6000 rpm (min. 1 rpm *)
Fluid connectionslip fittings with outside diameter 2 mm optional: manifold assembly M2.1
Wetted partsstainless steel 316L, ceramics, tungsten carbide Ni-based, epoxy resin; shaft seal: graphite-reinforced PTFE, stainless steel 316L;
static seals: FPM, optional: EPDM, FFPM
MotorDC-motor with graphite brushes, assigned power rating 4.5 W, nominal voltage 24 V, digital magnet encoder 32 counts per turn
Dimensions (L x W x H)Ø 18 mm, length 88 mm
Weightapprox. 100 g
Electrical connection10-pole connector
Differential pressure range0 – 5 bar (73 psi)
Maximum inlet pressure1 bar
Operating temperature range-20 … +60 °C
Displacement volume12 µl
Smallest dosage volume2 µl
Remark* depending on accessories,
Customized solutions on request

Even if single parameters are within the indicated performance range of technical data, certain parameter combinations may not be achievable. Single parameters may exceed their indicated performance range under adequate circumstances. For detailed evaluation please contact Thomson Process Equipment and Engineering who are the Irish Agent for HNP Mikrosysteme. Actual performance may vary.
Specifications are subject to change without notice.